Manufacturer:H-1043 Budapest, Dugonics u. 11.Tel.: (36-1) 889-0100 Fax: (36-1) 889-0200E-mail: @nivelco.com www.nivelco.comNIVELCO Process
10 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc DIMENSIONS HTK- - HTL- - HTV-- HTW- - HTR- - HTP- - HTS- - HTZ-
BKI 09 ATEX 0003X htk4014a0600p_01.doc 11 / 80 HTF- - HTG- - HTX- - HTY- - HTM- - HTQ- - HTI- -
12 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 1 Housing 2 Cable gland 3 High temp. connection 4 Mono cable probe 5 Weight 6 Twin cable probe 7
BKI 09 ATEX 0003X htk4014a0600p_01.doc 13 / 80 Nameplate 1 TYPE CODE AS ORDER CODE 2 Serial number Serial number 3 Manufacturing date 4 Amb
14 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 3.1 ACCESSORIES Certificate of Warranty PCStar2 software CD Installation and Programmin
BKI 09 ATEX 0003X htk4014a0600p_01.doc 15 / 80 4. MECHANICAL INSTALLATION 4.1 HANDLING AND STORAGE The device will weigh between approximately 3
16 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Avoiding blows - avoid hard blows, jolts, impacts, etc. Caution: fragile electronics Avoid bendi
BKI 09 ATEX 0003X htk4014a0600p_01.doc 17 / 80 Avoid cable kinks and fraying Do not coil the cable less than 400 mm or 16 “ in diameter. Cable kin
18 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 4.2 MOUNTING ON THE TANK 4.2.1 INSTALLATION INSTRUCTIONS: GENERAL NOTES Prior the installation s
BKI 09 ATEX 0003X htk4014a0600p_01.doc 19 / 80 Nozzles penetrating into tank Caution: Do not use nozzles that penetrate into the
2 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc GUIDED MICROWAVE LEVEL MEASUREMENT programmed measuring rangeof the applicationsfactory set valu
20 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Process connection and entry pipe Caution: Do not put the nozzle close to the entry pipe. Po
BKI 09 ATEX 0003X htk4014a0600p_01.doc 21 / 80 Probes: entanglement, straightness and tank bottom clearance Cable probes must be straight once i
22 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 1Agitator 2Support beam perpendicular to the pulse direction 3Abrupt changes in tank cross section
BKI 09 ATEX 0003X htk4014a0600p_01.doc 23 / 80 Fastening the probe to the tank bottom Flexible probes can be fastened with a chuck (ring), turnbu
24 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 4.2.2 SPECIFIC INSTALLATION INSTRUCTIONS: GAUGE - SOLID APPLICATIONS FALSE READINGS: 1 Do not
BKI 09 ATEX 0003X htk4014a0600p_01.doc 25 / 80 Conical silo nozzles, false readings and traction on the cable probes 2 High traction forces: W
26 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 4.3 WIRING Wiring in non-Ex environment 1 Detach the cover of the unit 2 Guide the cable into
BKI 09 ATEX 0003X htk4014a0600p_01.doc 27 / 80 Electrostatic discharge (E.S.D.) MicroTREK 2-wire non-Ex and Ex gauge electronics are shielded up
28 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Power supply Nominal voltage 24 V DC Maximum voltage (Uinput): 35 V DC Minimum voltage (Uinput):
BKI 09 ATEX 0003X htk4014a0600p_01.doc 29 / 80 Hazardous-duty version Connect the wires of the current loop to terminals 2 and 3 (any polarity).
BKI 09 ATEX 0003X htk4014a0600p_01.doc 3 / 80 CONTENTS 1. INTRODUCTION...
30 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 4.3.1 BUS ( HART®) COMMUNICATION Current out. and HART®, Passive, HART® protocol Microtrek has t
BKI 09 ATEX 0003X htk4014a0600p_01.doc 31 / 80 4.4 POWER-ON AND START-UP The MicroTREK 2-wire is pre-configured in accordance to customer order
32 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5. PROGRAMMING MicroTREK can be programmed in three (basic) ways. Programming with PCSTAR 2 sof
BKI 09 ATEX 0003X htk4014a0600p_01.doc 33 / 80 Installing the program: execute the file "PCSTAR.EXE" and follow the instructions on the
34 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Serial port – The serial port allows the user to select a free serial port (COM 1 to 4) on the co
BKI 09 ATEX 0003X htk4014a0600p_01.doc 35 / 80 After the connection has been established, the following screen is shown. This will show the curre
36 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 1.4.6 SUMMARY OF USER FUNCTIONS IN PCSTAR 2 (F2 – CONFIGURATION) The table below provides an over
BKI 09 ATEX 0003X htk4014a0600p_01.doc 37 / 80 Minimum 100 mm Maximum 24 000 mm 1.1.6 Probe length Ordered length + 100mm Set probe length
38 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Function Input Range Description 1.2.0 Display Select m, cm, mm, inch, nt h optional unit*
BKI 09 ATEX 0003X htk4014a0600p_01.doc 39 / 80 Function Input Range Description 1.3.0 Current output Select Off, Level, Distance, Volume*
4 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc
40 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Function Input Range Description 1.4 User data 1.4.3 Checksum Read only This value is us
BKI 09 ATEX 0003X htk4014a0600p_01.doc 41 / 80 Function Input Range Description 1.5.0 Application 1.5.1. Detection delay Minimum valu
42 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.1.3 QUICK CONFIGURATION: CONFIGURATION EXAMPLES The minimum functions (fct.) to be configured for
BKI 09 ATEX 0003X htk4014a0600p_01.doc 43 / 80 Typical gauge used for quick configuration examples: Probe type: twin Ø4 mm or 0.15” cable probe
44 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Example procedure 1 (using PCSTAR 2): to change true tank height (10000mm or 33 ft) to factory con
BKI 09 ATEX 0003X htk4014a0600p_01.doc 45 / 80 Dead zone: configuration of user menu PCSTAR 2 function 1.1.2 (HART® Fct. 2.1.1.4) The top dead z
46 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc How to set an analogue current output scale User menu functions 1.3.1 to 1.3.4 This set of functio
BKI 09 ATEX 0003X htk4014a0600p_01.doc 47 / 80 Setting up a volume table - (User menu PCSTAR 2 function 1.7.2 (HART® submenu 2.1.7)) To be able t
48 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Supplementary procedure for setting the 4 … 20 mA current to output volume readings (using PCSTAR
BKI 09 ATEX 0003X htk4014a0600p_01.doc 49 / 80 5.2 PROGRAMMING WITH SAP-300 DISPLAY UNIT The main parameters of the Microtrek can be also set u
BKI 09 ATEX 0003X htk4014a0600p_01.doc 5 / 80 Thank you for choosing a NIVELCO instrument. We are sure that you will be satisfied throughout its
50 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.2.2 MICROTREK’S BEHAVIOUR IN MANUAL PROGRAMMING MODE After power-up MicroTREK shows the measured
BKI 09 ATEX 0003X htk4014a0600p_01.doc 51 / 80 5.2.3 MANUAL PROGRAMMING Main menu Sub-menu PCSTAR 2 param. ID BASIC SETUP TANK HEIGHT 1.1.
52 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.3 PROGRAMMING WITH HART® HANDHELD (HHC) COMMUNICATOR Display and configuration can also be carr
BKI 09 ATEX 0003X htk4014a0600p_01.doc 53 / 80 Displaying readings Use the UP and DOWN arrows to move to the correct line and then select with th
54 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Configuration: summary of user functions via HART® Communicator HC 275 (Version 1.00) Reset defaul
BKI 09 ATEX 0003X htk4014a0600p_01.doc 55 / 80 Function (Fct.) Input Range Description 2.0 CONFIG./TEST 2.1.0 OPERATION 2.1.
56 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Function (Fct.) Input Range Description 2.1.2.1 LENGTH Select m, cm, mm, inch, Ft, optional
BKI 09 ATEX 0003X htk4014a0600p_01.doc 57 / 80 Function (Fct.) Input Range Description 2.1.3.0 ANALOG OUTPUT Select Level, Distance, Volume,
58 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Function (Fct.) Input Range Description 2.1.4.0 USER DATA 2.1.4.1 TAG 00000 01 Tag number
BKI 09 ATEX 0003X htk4014a0600p_01.doc 59 / 80 Function (Fct.) Input Range Description 2.1.5.0 APPLICATION 2.1.5.1.0 THRESHOLD 2.1.5.
6 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 2. ORDER CODE MicroTREK H - 4 - * 2-wire guided microwave level transmitter
60 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 2.1.7.0 STRAP TABLE Select m3, l, US Gal, Ft3, bbl, M3/h, Ft3/h, kg, Metric Tons, US Tons 2.
BKI 09 ATEX 0003X htk4014a0600p_01.doc 61 / 80 Function (Fct.) Input Range Description 5.0 HART® VARIABLES 5.1 MANUFACTURER NIVE
62 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.3.1 CHARACTERS AVAILABLE FOR ALPHA-NUMERICAL DATA FUNCTIONS IN PCSTAR2 AND ON THE HART® CONSOLE
BKI 09 ATEX 0003X htk4014a0600p_01.doc 63 / 80 5.4 MICROTREK 2-WIRE T.D.R. METER CHARACTERISTICS This subsection explains: the four principle
64 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc The configurations described below are illustrated in the above diagram (1) The “current output”
BKI 09 ATEX 0003X htk4014a0600p_01.doc 65 / 80 5.4.1 GAUGE OPERATING LOGIC WHEN THE REFLECTION IS LOST The product reflection pulse is usually l
66 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.4.2 GAIN AND VOLTAGE AMPLITUDE As explained in the measuring principle in the introduction, the
BKI 09 ATEX 0003X htk4014a0600p_01.doc 67 / 80 Level measurement: Level pulse amplitude and threshold After connection to a power supply, the Mi
68 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc The level signal can be optimized by way of two factors: Amplification factor The amplitude of the
BKI 09 ATEX 0003X htk4014a0600p_01.doc 69 / 80 1 Initial pulse 2 Flange reflection 3 Level signal 4 Interference signal 5 Threshold In the di
BKI 09 ATEX 0003X htk4014a0600p_01.doc 7 / 80 3. TECHNICAL DATA GENERAL DATA Measured values Between the reference point of the unit and reflect
70 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc 5.4.3 TYPICAL SIGNAL TRENDS The following diagrams show characteristic signals that have been rec
BKI 09 ATEX 0003X htk4014a0600p_01.doc 71 / 80 5.4.4 AUTOMATIC ADJUSTMENT To maintain a sufficiently strong reflection signal, the gain is adjust
72 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Procedure (example where the level measured is too low compared to true level): read off displayed
BKI 09 ATEX 0003X htk4014a0600p_01.doc 73 / 80 5.4.5 LEVEL MEASUREMENT WHEN MORE THAN ONE PHASE OR LAYER IN THE TANK Level can be measured with m
74 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Example application 2: level measurement when there are 2 liquids (oil/water) in the tank – using
BKI 09 ATEX 0003X htk4014a0600p_01.doc 75 / 80 5.5 TROUBLESHOOTING Event Fault Action Error messages “Tank full” status marker on*, reading fr
76 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc “Flange not found” status marker on* The signal converter has been incorrectly configured to measur
BKI 09 ATEX 0003X htk4014a0600p_01.doc 77 / 80 Event Fault Action General operation Instrument is not accurate with a product that has a high
78 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc Event Fault Action Electrical Connections and Communication Output No power supply Check the po
BKI 09 ATEX 0003X htk4014a0600p_01.doc 79 / 80 5.6 APPENDIX 1 – SET-UP PARAMETERS OF MICROTREK H-400 SERIES IN MULTICONT CONTROLLER Remote pro
8 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc TECHNICAL DATA OF THE PROBES TYPE HK - - HL - - HV - - HW - - HR--
80 / 80 BKI 09 ATEX 0003X htk4014a0600p_01.doc htk4014a0600p_01.doc April, 2009 Nivelco reserves the right to c