5.2.2 MICROTREK’S BEHAVIOUR IN MANUAL PROGRAMMING MODE
After power-up MicroTREK shows the measured value on the SAP display.
REMOTE MODE:
If the instrument senses external HART communication it changes its display mode and shows a "REMOTE MODE" message in the bottom of the display. In this
mode the measured values are refreshed according to the queries of the external HART master.
If the HART master does not refresh the display will show the last measured values.
In the absence of SAP-300 the COM LED indicates the HART communication.
If HART communication stops the COM LED turns off after 120 sec.
Entering the menu can be done by pressing the
E
button. Scrolling the menu can be done by pressing
and
buttons.
Enter into selected menu point with
E
button. Exit to the previous menu with button.
The buttons work only in presence of SAP-300 module.
When leaving the MicroTREK in (programming) menu after 30 minutes the instrument automatically returns to measuring mode.
If SAP-300 is removed the instrument instantly returns to measuring mode.
Because manual programming (with SAP-300) and remote programming (with external HART master, Multicont or PCSTAR2 software) cannot be performed at the
same time (as both of them act like a HART master) only one programming mode has priority and this is manual programming.
During manual programming the instrument sends the "BUSY" response to the external HART master device. (HART response code 32 - Device is busy)