BKI 09 ATEX 0003X
htk4014a0600p_01.doc
The configurations described below are illustrated in the above diagram
(1) The “current output” range is smaller than the max. possible measuring range
(2) The “current output” range is equal to the measuring range:
Scale min.: 4 mA (Fct. 1.3.3) = tank height – probe length + H
Scale max.: 20 mA (Fct. 1.3.4) = tank height – dead zone
(3) The “current output” range is greater than the measuring range:
Scale min.: 4 mA (Fct. 1.3.3) = 0.0
Scale max.: 20 mA (Fct. 1.3.4) = tank height – dead zone
(4) The “current output” range is greater than the measuring range:
Scale min.: 4 mA (Fct. 1.3.3) = tank height – probe length + H
Scale max.: 20 mA (Fct. 1.3.4) = tank height
(5) The “current output” range is greater than the measuring range:
Scale min.: 4 mA (Fct. 1.3.3) = 0.0
Scale max.: 20 mA (Fct. 1.3.4) = tank height
NOTE: The reference point for distance measurements is the bottom of the flange face.