BKI 09 ATEX 0003X
htk4014a0600p_01.doc
“Flange not found” status
marker on*
The signal converter has been
incorrectly configured to measure with a
cable or rod probe when it is equipped
with a coaxial probe. This may be also
due to installation on a long nozzle which
has the effect of attenuating the flange
pulse.
Contact NIVELCO for the corrective procedure.
“Delay out of limits” status
marker on*, reading is frozen.
The emitted pulse has not been
detected.
The gauge will not work until it has been
found.
The signal converter may need replacing. Please contact NIVELCO.
“Negative voltage error”*
“VC01 voltage error” *
“VC02 voltage error” *
“Reprogramming FPGA” *
Occurs when there is a problem with the
time base on the HF board.
Please contact NIVELCO.
*Gauge connected to PCSTAR2 software or HART Handheld Controller (HHC) with Marker status window (F8) or list displayed